imagen

Instruments


Wiener lab. CM 250
Wiener lab. CM 250

​Robust and user-friendly, Wiener lab. CM 250 operates with a leading edge software.

Throughput

  • Termostatized reagent dispenser

Sample Tray

  • 48 sample positions
  • Primary tubes or sample cups
  • Sample volume: 2 - 100 μL​
  • Continuous sample loading
  • Built-in barcode reader


Reagent Tray

  • 48 reagents positions
  • Peltier cooling system: up to 12ºC below Room Temperature (RT)l
  • 20, 30 and 50 mL vials​
  • Programmable volume:
    - R1: 0 - 700 μL
    - R2: 0 - 450 μL
  • Standard reagent volume: 200 μL

Dispenser  System​

  • Termostatized reagent dispenser
  • Liquid level detection by capacitive sensor level
  • Collision sensor
  • Inner and Outer probe washing system
  • Piston pump with valve assembly


Reaction Tray

  • 80 washable cuvettes with 0.6 cm light path
  • Heating system with air circulation (30ºC and 37ºC)


Washing System

  • Fully automatic 5-step washing and drying system
  • Water consumption: < 1 L/hour


Optical System

  • True double beam optical system
  • Interference filters: 340, 405, 505, 550, 570, 590, 650, 700 and 767
  • ​​Band width: 10 nm
  • Range: -0.1 - 3.6 Abs. (normalized to 1cm of light path)
  • Halogen lamp: 6V - 20W


Analytical Mode

  • Reagent or sample blank end point​
  • One and two-points kinetics
  • Factor, standard or calibration curve
  • Turbidimetry methods
  • Rutine, STAT, profiles
  • Automatic dilution of pathological samples or substrate consumption
  • QC: Levey-Jennings charts, Westgards rules
  • Samples pre-dilution if method requires


Data handling (requirements)

  • External PC - Printer
  • Import / export data, method and results
  • Software for Windows 10 (64 bits)
  • Automatic backup
  • Two avaible USB 2.0 ports


Communication

  • Serial port standard communication according ASTM 1394 protocol

​Installation Requirements

  • Power supply: 85 - 240 VAC, 43/65, 400 VA automatic set
  • Room Temperature: 15 - 35ºC; humidity: <80%; atmosphere: 750 - 1060 Hpa


​Dimensions

  • 80 cm (W); 48 cm (H); 57 cm (D)


Weight

  • Peso Bruto: 89 Kg
  • Peso Neto: 57 Kg


Certifications​

  • ISO 13485:2016, CE, ANMAT 3266/13
  • EMC: 61326-2-6-2006
  • ES: 61010-2-101:2002